Model Number N/A |
Device Problems
Break (1069); Unstable (1667); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report the cable break.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was inserted into the left atrium and the clip delivery system (cds) was advanced in the patient.The plus/minus knob was turned about a quarter when a click sound was heard, and the knob no longer worked.The cds and sgc were removed from the patient.Troubleshooting was performed outside the patient and the knob showed no resistance as it was continuously spinning.The physician suspected a cable break.A new sgc was selected and the same cds was used to complete the procedure reducing the final mr to a grade of 2.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
All available information was investigated and the reported cable break was confirmed via returned device analysis.The reported noise could not be replicated in a testing environment as it was likely a symptom of the cable break.The reported unstable/loose knob was not confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined the reported noise and unstable/loose knob appear to be related to the reported broken cable.However, a cause for how the cable broke cannot be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|