It was reported the catheter hub of a yueh centesis disposable catheter needle separated.The device required for a paracentesis to drain ascites and was placed via direct stick in the patient's abdomen.During the procedure, the hub of the catheter separated.No portion of the device was retained in the patient.It was noted that no force was exerted on the catheter.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation
a representative of covenant health care (united states) informed cook on 22feb2022 of an issue with a yueh centesis disposable catheter needle (rpn: (b)(4); lot:14382682).The customer stated the hub of the yueh broke off during a procedure.No adverse effects have been reported.No device was left in the patient.Reviews of the complaint history, device history record (dhr), drawing, quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One used device was received.The hub was broken/separated into two sections.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record (dhr) for lot 14382682.There are no nonconformances or additional complaints found for this lot.Cook did not identify any nonconforming material in house or in the field.Merit medical supplied the affected component.Cook requested they investigate this occurrence.The supplier reviewed one lot.The device history record showed the lot was released accomplishing all quality standard requirements.There was no deviation or nonconformance reported during the manufacturing of this product for this lot.It was confirmed that procedural and functional requirements needed for its release were met.The supplier has corrective action in place for this failure.Based on the dmr, dhr, device failure analysis, and supplier investigation, there is no evidence of manufacturing deficiency.Cook also reviewed product labeling.This complaint device is not currently packaged with instructions for use.Based on the information provided, inspection of the returned device, and the results of the investigation, cook concluded the cause of event could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|