SMITH & NEPHEW, INC. HEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72203697 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy, the healicoil anchor broke while being inserted into the pilot hole prepared.The pieces of the broken anchor were removed with tweezers.The procedure was completed with non-significant delay using a back-up device in the originally drilled bone hole.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A review of the anchor material drawing found storage conditions and composition test requirements with a specified expiration date for each lot.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details, the broken anchor was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Factors that may have contributed to the reported event include off-axial insertion, excessive force upon insertion, issues with bone hole preparation, or contact with another source.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|