| Model Number |
87163 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preparation for a procedure using a intellatip mifi xp catheter there was foreign material inside the sterile packaging.They replaced the catheter and were able to complete the procedure without patient complications.The catheter is expected to be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at boston scientific's post market laboratory the catheter and its sterile packaging was first visually inspected and the foreign material was found.The material was outside of the sterile pouch, stuck between the label and the exterior of the pouch.\the reported allegation was confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for a procedure using a intellatip mifi xp catheter there was a suspected foreign object contamination.They replaced the catheter and were able to complete the procedure without patient complications.The catheter was received for analysis.
|
| |
|