| Catalog Number |
11008552160 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
This report was initially provided to roche by fda via medwatch (b)(4).Roche urine dipstick products are giving trace protein results on a normal quality control.(expected result is negative).Chemstrip 7 (product number (b)(4) across multiple lot numbers, locations and users.Also observed in the chemstrip 10, and chemstrip 10sg product.Fda safety report id #:(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The customer dropped the qc solution onto the test pads, tipped the strip allowing the excess solution to run off the sides, and then read it.The recommended technique is to dip the strips into the qc solution.The retention material of lot 53644500, 55018600, and 51951500 were investigated and measured by visual reading.No false positive results were observed.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|