Catalog Number 11008552160
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
This report was initially provided to roche by fda via medwatch (b)(4).Roche urine dipstick products are giving trace protein results on a normal quality control.(expected result is negative).Chemstrip 7 (product number (b)(4) across multiple lot numbers, locations and users.Also observed in the chemstrip 10, and chemstrip 10sg product.Fda safety report id #:(b)(4).
 
Additional Manufacturer Narrative
The customer dropped the qc solution onto the test pads, tipped the strip allowing the excess solution to run off the sides, and then read it.The recommended technique is to dip the strips into the qc solution.The retention material of lot 53644500, 55018600, and 51951500 were investigated and measured by visual reading.No false positive results were observed.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHEMSTRIP 7
Common Device Name
URINE TEST STRIP
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14676398
Report Number1823260-2022-01716
Device Sequence Number12032945
Product Code JIL
UDI-Device Identifier00075537671603
UDI-Public00075537671603
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K032437
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number11008552160
Device Lot Number53644502
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-