• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. AMARA VIEW FULL FACE MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. AMARA VIEW FULL FACE MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 1090622
Medical Device Problem Codes Off-Label Use (1494); Use of Device Problem (1670); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 05/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer was made aware of an allegation by an end user that she developed "corneal erosion" in the past 6 months due to air blowing in her eyes during positive airway pressure (pap) therapy while using an amara view face mask.The user was reportedly prescribed unspecified eye drops and an ointment.The user has modified the mask by drilling holes in the bottom of it and refused the manufacturer's request that she notify her supplier.No part number or lot number have been provided.The manufacturer's investigation is ongoing.Upon completion of the investigation, a follow up report will be filed.
 
Additional Manufacturer Narrative
The manufacturer previously reported that an end user developed "corneal erosion" due to air blowing in her eyes during positive airway pressure (pap) therapy while using an amara view face mask.The user was reportedly prescribed unspecified eye drops and an ointment.The user modified the mask by drilling holes in the bottom of it and refused the manufacturer's request that she notify her supplier.The mask was returned to the manufacturer and the customer's complaint could not be confirmed.Though altered, the cushion was tested as received and did not leak.No airflow was detected in the eye area of the test setup mannequin.The investigation findings did not uncover any details that would change or modify the risk of the device.The product investigation laboratory observed three holes in the chin area of the cushion that were not of original design.Gauze tape was covering up the non-original holes.The rest of the cushion was damage and defect free and was of original design orientation.Due to the customer altering the cushion's original design, this product no longer meets philips respironics compliance and specifications.The amara full face mask is intended to provide an interface for application of cpap or bi-level therapy to patients.The mask is for single patient use in the home or multi-patient use in the hospital/institutional environment.The mask is to be used on patients (>66lbs/30kg) for whom cpap or bi-level therapy has been prescribed.This mask should also not be used on patients who need life support ventilation, who are taking prescription medication that may cause vomiting, or on patients who are uncooperative, obtunded, unresponsive, or unable to remove the mask.An exhalation port is built into the mask so a separate exhalation port is not required.Amara view minimal contact full face mask ifu provides instructions to check the mask and cushion before use as well as instructions on mask sizing.1109349 provides instructions for the user to discontinue use if eye irritation occurs.In addition, contraindications state that the mask may not be suitable for use on patients with recent eye surgery or dry eyes.The manufacturer could not determine the root cause for the complaint and concludes no further action is necessary.
 
Additional Manufacturer Narrative
Correction to the initial report 2518422-2022-42114 , section h-6 medical device problem code 3190 insufficient information was chosen.The correct selection is patient device interaction problem code-4001.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMARA VIEW FULL FACE MASK
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14679035
Report Number2518422-2022-42114
Device Sequence Number20180485
Product Code BZD
UDI-Device Identifier00606959009196
UDI-Public00606959009196
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082866
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1090622
Device Catalogue Number1090622
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer08/15/2022
08/15/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date08/17/2022
12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-