Catalog Number 8065752554
Medical Device Problem Codes Fitting Problem (2183); Defective Component (2292); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during a surgery an ophthalmic laser probe was stuck, the tip was found damaged and the light beam was not focusing correctly.Procedure was completed by replacing the product.There was no patient impact.
 
Additional Manufacturer Narrative
The 23-gauge articulating illuminated ophthalmic laser probe was received and a visual assessment of the returned sample revealed the sample to have a damaged tip.However, as to how or when it became damaged remains inconclusive.Trocar testing was performed and revealed excess adhesive on cannula-nitinol junction.The reported event of the stuck laser probe was confirmed.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.A potentially relevant complaint was found and reviewed as part of this investigation.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The root cause of the reported event of the stuck laser probe is attributed to excess adhesive on cannula-nitinol junction.Based on the information obtained, the root cause of the reported event of tip damage is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASER PROBES
Common Device Name
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14679253
Report Number2028159-2022-00763
Device Sequence Number19992517
Product Code HQE
UDI-Device Identifier00380657525546
UDI-Public00380657525546
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K946135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number8065752554
Device Lot Number13L9J5
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date09/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-