Model Number Q6 EDGE HD
Medical Device Problem Codes No Apparent Adverse Event (3189); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 05/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event or Problem Description
Provider alleges consumer alleges the chair went into bluetooth and turned on its own into the wall where the consumers foot allegedly got stuck.
 
Additional Manufacturer Narrative
The alleged performance condition could not be duplicated.The device's drive function is not operable in bluetooth mode.The device will go into bluetooth mode when a controller error occurs but the drive function is disabled.There were no error codes in the log for the reported date of incident.(b)(6) 2022).The device's appearance indicates the device was subjected to a great deal of abuse/misuse.Despite this, the alleged performance anomaly could not be duplicated during functional test.
 
Event or Problem Description
Provider alleges consumer alleges the chair went into bluetooth and turned on its own into the wall where the consumers foot allegedly got stuck.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
POWERED WHEELCHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key14679557
Report Number2530130-2022-00051
Device Sequence Number14666574
Product Code ITI
UDI-Device Identifier00606509200059
UDI-Public00606509200059
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K143383
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberQ6 EDGE HD
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexMale
-
-