COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET Back to Search Results
Catalog Number IGTCFS-65-FEM
Device Problem Structural Problem (2506)
Patient Problems Abdominal Pain (1685); Perforation of Vessels (2135)
Event Type  Injury  
Manufacturer Narrative
The previous medwatch report was submitted by william cook europe under manufacturer report reference# 3002808486-2022-00075.Additional information provided determined that this device was manufactured by cook inc.With the submission of this initial medwatch report, cook inc.Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # referenced in this initial medwatch report.Occupation: non-healthcare professional investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava perforation, abdominal pain, limited physical activity.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported abdominal pain and limited physical activity is directly related to the filter and unable to identify a corresponding failure mode at this point in time.A total of 20 devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event Description
Patient allegedly received a gunther tulip inferior vena cava (ivc) filter on (b)(6) 2009 via the left common femoral vein due to conjunction with an orthopedic procedure.Patient is alleging vena cava perforation.Patient alleges "constant abdominal pain." the patient is limited or can no longer engage in "bending over, lifting weights".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP FEMORAL VENA CAVA FILTER SET
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key14680588
MDR Text Key294880983
Report Number1820334-2022-01049
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043509
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2012
Device Catalogue NumberIGTCFS-65-FEM
Device Lot Number2272665
Was Device Available for Evaluation? No
Date Manufacturer Received05/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
Patient Weight147 KG
-
-