| Model Number |
PC-3-8-3D |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information reporting that three axium coils had significant resistance in the sheath when trying to advance into the catheter hub due to robbin-tapered coil sheath.It was noted that the coils and all accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or injury were reported.Ancillary device: stryker sl-10 microcatheter additional information received reported that there was no malformation or suspected defect of the coil sheath.The gap between the coiling sheath and hub caused the coil to break before entering the coiling catheter causing the heavy resistance.The coil never entered into the coiling catheter or patient.There were no complications as they were not used with the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported all three coils encountered heavy resistance at level of distal coil sheath and hub of micr ocatheter.Resistance at gap produced by non tapered coiling sheath and hub of microcatheter.No other issues noted by physician.
|
| |
|