| Model Number |
21-6500-13 |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during the use of the product, the patient felt there was an anomaly in the infusion rate.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.The customer stated problem was not duplicated.No device problem was found.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.The device was manufactured in apr 2015, and is out of warranty.Therefore, no manufacturing dhr review is needed.
|
| |
|