• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD-LEGACY PLUS AMBULATORY INFUSION SYSTEM; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD-LEGACY PLUS AMBULATORY INFUSION SYSTEM; PUMP, INFUSION Back to Search Results
Model Number 21-6500-13
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that during the use of the product, "no disposable, pump won't run" alarm went off.No patient injury was reported.
 
Additional Manufacturer Narrative
H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.The customer stated problem was not duplicated.As a result of checking the log, it was confirmed that there was a record of repeated power on/off on may 18, 2022, presumably caused by a cassette recognition failure.Replaced the downstream occlusion sensor and the upstream occlusion sensor was re-calibrated.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.The device was manufactured jun.2015, and is out of warranty.Therefore, no manufacturing dhr review is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY PLUS AMBULATORY INFUSION SYSTEM
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key14682136
Report Number3012307300-2022-11750
Device Sequence Number14857302
Product Code FRN
UDI-Device Identifier10610586022593
UDI-Public10610586022593
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051568
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number21-6500-13
Device Catalogue Number21-6500-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-