• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. WEREWOLF ENT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. WEREWOLF ENT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 72290144
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a turbinate reduction procedure, the adapter of the werewolf ent started smoking and burned.It is not known if the problem is the adaptor or the console.The procedure was successfully completed with a non-significant surgical delay using a back-up device.No patient complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection observed the warranty seal is broken.No burn/arc marks are seen in ether wand port.A functional evaluation revealed no issues.The unit was opened and found damage around the power supply from removal of the part, the railing is heavily bent, scratches all around the railing, and a rivet is found loose in the unit.No burn or arc damage was found.It was determined the device did not contribute to the reported event.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that could have contributed to the reported event include a failure of a concomitant device.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reported events.A review of risk management files found no updates to the risk files are recommended at this time.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF ENT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14682169
Report Number3006524618-2022-00291
Device Sequence Number5503214
Product Code GEI
UDI-Device Identifier00885556739624
UDI-Public00885556739624
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K202006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72290144
Device Catalogue Number72290144
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1