Model Number U128 |
Device Problems
False Positive Result (1227); Over-Sensing (1438); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that this implantable pacemaker recorded a signal automatic monitoring (sam) episode due to an intrinsic atrial fibrillation episode experienced by the patient.Due to this, the minute ventilation (mv) sensor was disabled.Furthermore, low out of range pacing impedance measurements were exhibited from both the right ventricular and right atrial channels.Troubleshooting options were discussed.This device remains in service.No further adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Additional information was added to the following fields: h6: device codes.
|
|
Event Description
|
It was reported that this implantable pacemaker recorded a signal automatic monitoring (sam) episode due to an intrinsic atrial fibrillation episode experienced by the patient.Due to this, the minute ventilation (mv) sensor was disabled.Furthermore, low out of range pacing impedance measurements were exhibited from both the right ventricular and right atrial channels.Troubleshooting options were discussed.This device remains in service.No further adverse patient effects were reported.
|
|
Search Alerts/Recalls
|