| Model Number |
1309 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Tooth Fracture (2428)
|
| Date of Event |
05/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The scan sent by the doctor to prosomnus seemed to be distored as the device was designed around the files received.The finished device passed the prosomnus quality standard at final quality control.The doctor also mentioned that they extensively ground the arch to try making it fit which is another indication that the records received were not accurate.The device broke during the adjustment due to the extent of the distortion.The patient had a crown made to replace the chipped tooth and a new device has been made on the new scan received.
|
| |
|
Event or Problem Description
|
|
Patient's tooth #7 chipped off during her delivery appointment.The doctor reported that she tried to adjust the tray, but it broke in half in the #7 area.
|
| |
|