Model Number CD2257-40Q
Medical Device Problem Codes Failure to Capture (1081); Incorrect Measurement (1383); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2022-12161.It was reported that a patient presented in-clinic for an explant procedure.Examination of the patient's implantable cardioverter defibrillator (icd)was found to have exhibited a diagnostic anomaly, as the device was not able to return a high voltage impedance reading.Additional malfunctions of low pacing impedance and loss of capture were noted on the device.The patient's right ventricular (rv) lead was found to have malfunctions of high capture thresholds and insulation break.The rv lead was capped and the icd was explanted on (b)(6) 2022.Both the icd and rv lead were replaced.No patient consequences were reported.
 
Additional Manufacturer Narrative
The reported event of capture anomaly and impedance anomaly could not be confirmed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.A longevity calculation was performed and found the battery depletion was normal based on the device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD2257-40Q FORTIFY ASSURA DR ICD_IDE_US
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14683214
Report Number2017865-2022-12162
Device Sequence Number4675115
Product Code NVZ
UDI-Device Identifier05414734505550
UDI-Public05414734505550
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2014
Device Model NumberCD2257-40Q
Device Catalogue NumberCD2257-40Q
Device Lot Number4132788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-