ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 04/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lens was returned wrapped in paper.Solution was dried on the lens.The optic was cut into pieces.The cut optic pieces had multiple scratches and split/cracks on the anterior and posterior sides of the lens.The observed damage is typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The file indicated the use of a non-qualified viscoelastic.The product investigation could not identify a root cause for the reported complaint.Power and resolution testing could not be conducted due to the extensive optic damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.There are no other complaints in this lot.Investigation has been completed based on current information.Based on our current tracking, there are no adverse trends for this reported complaint.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported following the intraocular lens implantation the patient was slightly unhappy with the vision.The lens was exchanged in a secondary procedure.The clinical reason for explant was "slightly myopic".Additional information has been received that, there is no patient harm involved.The patient was not hospitalized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14683269
Report Number1119421-2022-01255
Device Sequence Number5083908
Product Code POE
UDI-Device Identifier00380652424677
UDI-Public00380652424677
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/16/2025
Device Model NumberDAT015
Device Lot Number15099704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL CARTRIDGE C; MONARCH III IOL CARTRIDGE D; NUVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR