| Model Number |
MA50BM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Corneal Edema (1791); Capsular Bag Tear (2639)
|
| Date of Event |
04/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non health care professional reported that, during an intraocular lens (iol), it was found that haptic broken off.The surgeon had to cut lens in 2 pieces to remove from eye during initial procedure.Additional information has been received and stated that, there was a posterior capsular tear for which a 3-piece iol was placed in the sulcus.Since there was a risk of iol drop to the vitreum, the hind leg was held back, this leg came loose and sulcus lens had to be cut in anterior chamber to be removed.The patient was improving from the corneal edema.The patient was treated with hyperosmolar ophthalmological solution and ointment.There were no other complications to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information provided in b.7.On initial mdr submission posterior capsule tear was reported as an adverse event but the tear was happened before injecting the intraocular lens.Additional information provided in b.7., h.3., h.6., h.10.A non-qualified cartridge was indicated.The company lens model is only qualified for use in the company cartridge.The root cause to for the reported haptic damage may be related to a failure to follow the dfu (directions for use).The company lens is only qualified for use in the company cartridge.The ifu (instructions for use) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company foldable iols (intraocular lens) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.In addition, information was provided that since there was a risk of the lens falling into the vitreous the trailing haptic was held back.It is unknown with what and how the haptic was "held back".This additional manipulation may have cause the reported damage.Haptic manufacturing is a validated process with multiple inspections to verify the quality of the produced haptic material.These inspections include a verification of the dimensional accuracy, tensile strength, flexibility and cosmetic appearance.The inspections ensure that the material meets all required specifications before it is allocated for use.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|