• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATOS MEDICAL AB PROVOX NID 20FR, 6MM; INDWELLING VOICE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATOS MEDICAL AB PROVOX NID 20FR, 6MM; INDWELLING VOICE PROSTHESIS Back to Search Results
Model Number 20 Fr
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Pneumonia (2011)
Date of Event 04/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report includes a theoretical investigation since no product was returned to the manufacturer.According to the information given in the complaint form.Serious injury: no.Medical intervention required: yes.Early leakage of a prosthesis is mostly caused by food remnants or mucus around the flap towards the sealing surface of the valve seat.The ifu gives the direction to clean the prosthesis thoroughly and regularly, morning and night and after every meal.It is not possible to state why the prosthesis was leaking without examining it.During the manufacturing of a nid prosthesis, there are several visual checks.Leakage- and pressure drop tests are also performed.It is not likely that the prosthesis was leaking when leaving the manufacturing unit at atos medical.The prosthesis is not available for investigation, however, the most likely explanation to the leakage is user error.This conclusion is based on: early leakage is mostly due to stuck food or mucus between the flap and the sealing surface of the blue ring.The ifu states that cleaning should be done after every meal.There was only leakage during one day before the prosthesis was replaced.When replacing, the clinician must check for leakage according to ifu.It is not likely that the prosthesis was damaged when leaving the manufacturing unit at atos medical.Leakage is normally leading to coughing which is an indicator for replacement of the prosthesis.(b)(6) has been consulted and they are stating that pneumonia may develop, but not likely, within 24h if the leakage is severe and the patient has very poor or absent cough reflex.Atos medical does not know if the leakage was severe or if the patient is able to cough properly or if the cough reflex was reduced or absent (there was a previous stay at the hospital which cause is not presented in the complaint description).Conclusion: root cause cannot be unambiguously determined as some facts are missing.However, given that the patient was treated with antibiotics (medical intervention), we prefer to report this as-is to err on the side of caution even if it is difficult to confirm that the pneumonia was a result of a leaking prosthesis that was used for only one day.Note: due to confidentiality reasons, the patient's name is not entered as the initial reporter.Instead, the sales representative's name is used.
 
Event or Problem Description
This is the event description received from the initial reporter: patient was in hospital for about 1.5 weeks, came home, had additional problems so had to go back in.It turned out the valve was leaking and caused aspiration pneumonia in which she had to be put on antibiotics.Several follow-up questions have been asked and were answered by the customer support representative since it was not possible to reach the user.The user only had the prosthesis for one day.The user went to hospital for other reasons and the doctor diagnosed pneumonia due to the leaking prosthesis.The nid was replaced with another vp so that a plug can be used.The user is out of the hospital and has ordered another nid, which the user is normally using.Description of the product: provox nid is a non-indwelling voice prosthesis for patients who are capable of handling the exchange and maintenance of a voice prosthesis independently of a clinician or physician.The prosthesis is available in two outer shaft diameters (17 and 20 french) and several lengths.The voice prosthesis is seated in a puncture in the wall between esophagus and trachea.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVOX NID 20FR, 6MM
Common Device Name
INDWELLING VOICE PROSTHESIS
Manufacturer (Section D)
ATOS MEDICAL AB
kraftgatan 8
hörby, 24235
SW  24235
Manufacturer (Section G)
ATOS MEDICAL AB
kraftgatan 8
hörby, 24235
SW   24235
Manufacturer Contact
karolina nilsson
kraftgatan 8
hörby, 24235
SW   24235
MDR Report Key14684371
Report Number8032044-2022-00005
Device Sequence Number14621418
Product Code EWL
UDI-Device Identifier07331791000331
UDI-Public07331791000331
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K043138
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number20 Fr
Device Catalogue Number7111
Device Lot Number2201087
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-