Model Number LOGIC FEMORAL PS CEM LEFT SZ 2
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately seven years post initial left tka, the 65 y/o female patient started having pain.Surgeon had removed a loose femoral implant that need revised to a size 2 condylar constraint femur, along with replacing the poly to a 17mm constrain liner.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.X-rays and photos received.The devices are available for evaluation.
 
Additional Manufacturer Narrative
Concomitant medical products: logic ps tibial insert size 2 13mm (cat# 02-012-35-2013 / serial# (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the revision reported was likely the result of both patient-related conditions and an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.The loosening likely led to gross instability and allowed the femoral component to migrate into flexion and internally rotate, which caused an increase in cement and bone particles in the joint space and resulted in tibial insert wear (articulating surface and backside) and tibial insert subsidence.The most probable root cause associated with the reported event of ¿loosening - femoral¿ is associated with weakened integration of the femoral component at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.Furthermore, the most probable root cause associated with the event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 2
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14684530
Report Number1038671-2022-00680
Device Sequence Number5538829
Product Code JWH
UDI-Device Identifier10885862001122
UDI-Public10885862001122
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2019
Device Model NumberLOGIC FEMORAL PS CEM LEFT SZ 2
Device Catalogue Number02-010-01-0220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer07/11/2023
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date08/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight104 KG
-
-