| Model Number |
60-5274-032 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Conmed (b)(4) reported on behalf of their customer that the device, 60-5274-032, spatula electrod w/stealth er was being used on 30may22 during a laparoscopic cholecystectomy procedure when it was reported ¿that the tip of electrode was detached during the surgery.It was dropped off into the patient body but it was successfully retrieved from the body.¿ an alternate unknown device was used to complete the surgery." there was no impact or injury to the patient or user.There was no report of medical intervention or any prolonged hospitalization for the patient.After further assessment it was found that there was a delay of ¿several minutes¿.The fragment was retrieved with laparoscopic forceps.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.Device evaluated by mfr: device not yet received.
|
| |
|
Event or Problem Description
|
|
Conmed japan reported on behalf of their customer that the device, 60-5274-032, spatula electrod w/stealth er was being used on (b)(6) 2022 during a laparoscopic cholecystectomy procedure when it was reported ¿that the tip of electrode was detached during the surgery.It was dropped off into the patient body but it was successfully retrieved from the body.¿ an alternate unknown device was used to complete the surgery." there was no impact or injury to the patient or user.There was no report of medical intervention or any prolonged hospitalization for the patient.After further assessment it was found that there was a delay of ¿several minutes¿.The fragment was retrieved with laparoscopic forceps.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Received one 60-5274-032 in opened original packaging.Lot number was verified.Performed a visual inspection, the spatula electrode is disassembled from the device.There is evidence of soldering present on the electrode.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised the following: examine the device prior to use for damage.Do not use if damage is found.These electrosurgical suction/irrigation instruments should only be used by surgeons trained in surgical endoscopy according to established hospital protocol.To avoid potential damage to trocar seals and to the electrosurgical tip, always slide the tip shield in the covered and locked position during insertion into the trocar cannula.Misuse of this or any other electrosurgical device can result in a patient injury.Read and become familiar with all instructions and directions before using this device.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|