Model Number SN60WF
Medical Device Problem Codes Break (1069); Inaccurate Delivery (2339)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 05/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during the cataract surgery with intraocular lens implant procedure, the procedure was not gone well and they put the lens back in the cartridge.On the second occasion the lens was implanted and rotated correctly.Three days following the procedure the patients visual acuity did not improve.When checked in the lamp of slit, the haptic was broken for which the lens needs to be removed.During the surgery it was not observed, in fact the lens was rotated and it did not look broken.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified associated products were used.The product investigation could not identify a root cause for the reported complaint.The product was not returned.It is unknown if qualified associated products were used.The instructions for use (ifu) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Information was provided that the lens was loaded and delivered twice.Lenses and cartridges are singe-use devices.The broken haptic may have occurred due to the additional manipulation/loading issues.No further information has been provided.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14685049
Report Number1119421-2022-01263
Device Sequence Number5617396
Product Code HQL
UDI-Device Identifier00380655093306
UDI-Public00380655093306
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/22/2025
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.255
Device Lot Number15122655
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer07/30/2022
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-