Model Number 26055CB
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As per manufacturer incident report #(b)(4) received from the factory in germany: the reported event happened in spain.According to the information available, the ceramic tip broke off inside the patient, with a kind of explosion.The physician had to remove the parts from inside the patient without being sure that all parts were collected.There is no further information related to the described explosion.
 
Additional Manufacturer Narrative
The initial review of the instruments received for analysis does not show signs of explosion; thermal damages and abrasion are visible on the instruments.Investigation is ongoing.
 
Additional Manufacturer Narrative
The explosion/bang described by the user can only have been caused by the damaged ceramic insert.The investigation showed that the ceramic was damaged by excessive force application.In the course of the missing insulation between the optics and the inner shaft surface, a short circuit subsequently occurred.The distal end shows thermal damage and the ceramic insert is completely broken due to excessive force application.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH F. RESECT. SHEATH, 15 FR.
Common Device Name
RIGID ENDOSCOPE SHEATH, REUSABLE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM  
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key14685225
Report Number9610617-2022-00090
Device Sequence Number6636957
Product Code HIH
UDI-Device Identifier04048551275597
UDI-Public4048551275597
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K882270
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26055CB
Device Catalogue Number26055CB
Device Lot NumberRO01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date07/12/2022
Date Device Manufactured04/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-