Model Number IPN000864
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).A visual, dimensional, nor functional inspection of the product involved in the complaint could not be conducted since the product was not returned.No issues or discrepancies were found in the device history record of product code 833-137 / batch 74l2102767 that can be related to the failure mode reported.Incoming inspection records of the needle lot involved in the manufacturing of product code 163625ph / batch 3200649200 were reviewed, no issues or discrepancies were found.Also, vendor record of the suture lot involved in the manufacturing of product code dmmv100/batch m210728vb2-300r were reviewed, no issues or discrepancies were found.A corrective action cannot be implemented at the time since the sample involved in the customer complaint was not sent for analysis.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility is not being manufactured at the time.If the sample becomes available this investigation will be updated with the evaluation results.
 
Event or Problem Description
Medwatch report# (b)(4).Dart portion at the end of suture broke off and resulted in a retained foreign body.Disclosed to family and patient and decision to leave in place was made.Another similar incident occurred but the dart was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MDK VI MF 0 TC43/HR26 48"
Common Device Name
SUTURE, SURGICAL, ABSORBABLE,
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key14685379
Report Number1036844-2022-00042
Device Sequence Number17495451
Product Code NEW
UDI-Device Identifier14026704603920
UDI-Public14026704603920
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K030212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2024
Device Model NumberIPN000864
Device Catalogue Number833-137
Device Lot Number74L2102767
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Initial Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-