• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-172-C
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 06/04/2022
Event Type  malfunction  
Event Description
A report was received on (b)(6) 2022 from the home therapy nurse (htn) of a 57-year-old male patient with a medical history including end stage renal disease, who stated the patient experienced a cartridge blood leak resulting in an approximate 500-800ml blood loss during a home hemodialysis training on (b)(6) 2022.The patient attended the outpatient room of hospital.Additional information was received on (b)(6) 2022 from the htn stating the patient experienced increased blood pressure (nos) and was assessed at hospital with no laboratory changes or medical intervention.Per the htn, the patient continues to treat using the nxstage system.
 
Manufacturer Narrative
The cartridge was discarded and not received for evaluation.All product is released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "check the system for blood and fluid leaks during treatment, and pay close attention to the blood line and access connections".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
MDR Report Key14686220
MDR Text Key301750719
Report Number3003464075-2022-00027
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K140526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCAR-172-C
Device Catalogue NumberCAR-172-C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/04/2022
Initial Date FDA Received06/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age57 YR
Patient SexMale
Patient Weight79 KG
-
-