• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER2 STRAIGHT ROD-480MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER2 STRAIGHT ROD-480MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 186789480
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional product codes: osh,kwp,mnh,nkb and mni.Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthese reports an event in japan as follows: it was reported that on (b)(6) 2018 the patient underwent a primary spinal fusion procedure.On (b)(6) 2022 the patient underwent a removal procedure because the fusion in the lesion was obtained.Prior to the procedure, it was reported that the rod (t10-l3) had broken off.The removal procedure was completed, and all the devices were removed.The rod¿s breakage did not caused adverse effect to the patient.Concomitant device reported: unknown locking/set screws (part# unknown; lot# unknown; quantity: 1).This report is for one (1) viper2 straight rod-480mm this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the photos.Visual analysis of the post operative x-rays and the product photos revealed that the reported viper2 straight rod-480mm broke in two fragments.No other problems were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for viper2 straight rod-480mm.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a review of the receiving inspection (ri) for viper2 straight rod-480mm was conducted identifying that lot number tbtkd was released in two batches; and identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER2 STRAIGHT ROD-480MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 02400
SZ   02400
MDR Report Key14688974
Report Number1526439-2022-00837
Device Sequence Number5465239
Product Code KWQ
UDI-Device Identifier10705034133522
UDI-Public(01)10705034133522
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111136
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186789480
Device Catalogue Number186789480
Device Lot NumberTBTKD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/22/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-