| Model Number |
188.044 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient's spinous process broke post-operatively.Subsequently, sp-fix implant has migrated.This event occurred in germany.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the device or any imaging could be provided for evaluation.No determinations could be made as to the cause of the reported issue.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient's spinous process broke post-operatively.Subsequently, sp-fix implant has migrated.This event occurred in germany.
|
| |
|
Additional Manufacturer Narrative
|
|
The peek barrel was returned, connected to two plates.Additional information provided that the surgeon believes he may have used a barrel too large for the patient's anatomy.It is possible that the used of incorrect barrel size in conjunction with poor bone quality may have caused the spinous to break off.
|
| |
|