| Model Number |
133S-MV-16-T |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: broken device.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported unknown side one of ¿allergan¿s tissue expanders had a hole in it.If i got the story correct it was implanted before they discovered the problem¿.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported unknown side one of ¿allergan¿s tissue expanders had a hole in it.If i got the story correct it was implanted before they discovered the problem¿.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported unknown side one of ¿allergan¿s tissue expanders had a hole in it.If i got the story correct it was implanted before they discovered the problem¿.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: two openings on anterior.Leak test and microscopic analysis was performed which identified: two striated openings on anterior and non-penetrating nick striated on anterior.Based on the device analysis the final assessment is: two striated openings on anterior assessed as surgical damage.
|
| |
|