MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY FORCEPS Back to Search Results
Model Number FB-19K-1
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During inspection and testing, foreign matter was found on the arms of the cups (the linkage mechanism at the distal end) of the forceps.A review of the device history record found no deviations that could have caused or contributed to the foreign material being adhered.There were no anomalies noted in the operation or appearance of the tip.Based on the results of the legal manufacturer's investigation, it is likely the adhesion of foreign material is due to insufficient cleaning, or that it has not been sufficiently cleaned due to the passage of time after use.However, a definitive root cause of the reported issue could not be identified.The device's instruction manual provides the following warning, which may help to prevent the issue: "reprocess the instrument immediately after use by immersing it in a neutral, low-foaming, medical-grade detergent solution, then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off during use and may interfere with operation of the cups"."when inserting the instrument into the endoscope, hold it close to the biopsy valve or t-plug and keep it as straight as possible relative to the biopsy valve or t-plug.Otherwise, the insertion portion could be damaged".Olympus will continue to monitor field performance for this device.The report of there being a catch when operating the slider to open and close the cup was confirmed.There was a feeling of the forceps being caught when opening.After lubricant was applied, the cup opened and closed smoothly.Per the legal manufacturer, this device defect has no potential to cause or contribute to death or serious injury if the malfunction were to recur.
 
Event or Problem Description
The customer reported that, during preparation for use, there was a catch when the slider was moved back and forth during the pre-case inspection.The cup opened and closed, but there was a catch.The procedure was completed with another similar device.The subject device was sent to an olympus service center for evaluation.During inspection and testing, foreign matter was found on the forceps.This report is being submitted for the malfunction found during evaluation (foreign material).There was no harm or user injury reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPSY FORCEPS
Common Device Name
BIOPSY FORCEPS
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14690003
Report Number8010047-2022-10022
Device Sequence Number15112595
Product Code EOQ
UDI-Device Identifier04953170029608
UDI-Public04953170029608
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K962555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFB-19K-1
Device Lot NumberK1Y17-1757
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse