CVRX, INC. BAROSTIM NEO; CAROTID SINUS LEAD Back to Search Results
Model Number 2102
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Hematoma (1884)
Date of Event 05/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received from the site, the hematoma could have been related to the anticoagulant medication the patient was taking; however, the cause of the infection was unable to be undetermined.The device history and sterilization records for this device serial number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
A barostim system was implanted on (b)(6) 2022.The patient was admitted to the hospital on (b)(6) 2022 due to hematoma at the pocket site.It was noted that the patient was taking an anticoagulant medication.An infection culture was performed, and proteus mirabilis was detected.The hematoma site was evacuated, and moderate old liquified hematoma was observed with no gross signs of infection.The pocket was irrigated with saline with a flow seal, the ipg was placed in an absorbable antibiotic pouch, and oral antibiotics were administered.Therapy remained activated, and the patient experienced functional improvement in their heart failure.The patient was discharged home on (b)(6) 2022 with the infection controlled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAROSTIM NEO
Common Device Name
CAROTID SINUS LEAD
Manufacturer (Section D)
CVRX, INC.
9201 west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer (Section G)
CVRX, INC.
9201 west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer Contact
sarah hicks
9201 west broadway avenue
suite 650
minneapolis, MN 55445
MDR Report Key14690043
Report Number3007972010-2022-00008
Device Sequence Number17815838
Product Code DSR
UDI-Device Identifier00859144004432
UDI-Public(01)00859144004432(17)231006
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/06/2023
Device Model Number2102
Device Catalogue Number100054-202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Initial Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexMale
Patient Weight115 KG
Patient RaceBlack Or African American
-
-