| Model Number |
1020-74-220S |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
|
| Health Effect - Clinical Codes |
Drug Resistant Bacterial Infection (4553); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional procode: kwp.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: on (b)(6) 2021, a patient underwent a primary posterior cervical fusion in the occipital bone and the procedure was completed without surgical delay.On (b)(6) 2022, the surgeon reported that the rod might have come off, and a revision procedure was scheduled for (b)(6) 2022.It was reported that during the revision procedure on (b)(6) 2022, the surgeon commented that the setscrews on both ends of the plate had come off, and that the rod was about to protrude from the occipital region.This report is for a symphony oct system adjustable rod titanium alloy diameter 4.0 to 4.0x220mm.This is report 2 of 2 for (b)(4).Additional reports are captured under (b)(4).
|
| |
|
Event or Problem Description
|
|
Mrsa infection is confirmed.The surgeon commented that the patient¿s son is on the verge of suing for lack of explanation from the manufacturer about implant problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b5: event description updated.D3: manufacturer email address added.E1: initial reporter's phone number added.A manufacturing record evaluation was performed for the finished device product code: 102074220s.Lot number: rl287627.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 13.10.2020.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|