• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; CANAL-FILLING STEM EXTENSION, 12X100MM, STRAIGHT, SPLINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; CANAL-FILLING STEM EXTENSION, 12X100MM, STRAIGHT, SPLINED Back to Search Results
Model Number KSP11100E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00186, #3013450937-2022-00187, #3013450937-2022-00188, #3013450937-2022-00189, #3013450937-2022-00190, #3013450937-2022-00191.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: tibial poly spacer, tibial hinge component, distal femur, and distal femur axial pin.The following eleos implants remain implanted from the patient's original surgery: cemented segmental stem, canal-filling stem extension, and tibial baseplate.No additional information regarding this adverse event has been reported.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: tibial poly spacer, tibial hinge component, distal femur, and distal femur axial pin.The following eleos implants remain implanted from the patient's original surgery: cemented segmental stem, canal-filling stem extension, and tibial baseplate.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added.G3: date received by manufacturer added.G6: type of report added.H2: follow-up type added.H3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.H10: additional narratives/data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
CANAL-FILLING STEM EXTENSION, 12X100MM, STRAIGHT, SPLINED
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
MICROPORT ORTHOPEDICS
5677 airline road
arlington TN 38002
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key14690807
Report Number3013450937-2022-00192
Device Sequence Number12457083
Product Code KRO
UDI-Device IdentifierB278KSP11100E0
UDI-PublicB278KSP11100E0
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2025
Device Model NumberKSP11100E
Device Catalogue NumberKSP11100E
Device Lot Number1716449
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date06/14/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/19/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR.; P/N 25000011E, ELEOS CEMENTED SEGMENTAL STEM.; P/N 25001212E, ELEOS POLY SPACER.; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN.; P/N 25002204E, ELEOS TIBIAL BASEPLATE.; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-