| Model Number |
5303-90004 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tip of an aleutian an coronal inserter was fractured and the outer shaft/handle was broken outside the or.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the handle.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, functional and material analysis could not be performed as the device was not returned.Complaint history records were reviewed for this catalog number, similar complaints were identified.The device was not returned and is unavailable for evaluation.Photos provided show an intact outer shaft, therefore, it is likely that there is no problem with the outer shaft, and the reported fracture is only affecting the inserter.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of an aleutian an coronal inserter was fractured and the outer shaft/handle was broken outside the or.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the handle.
|
| |
|