• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR TIB INSERT SLOPE +, SZ 2.5, 9 MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR TIB INSERT SLOPE +, SZ 2.5, 9 MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number LOGIC CR TIB INSERT SLOPE +, SZ 2.5, 9 MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 4.3 years post initial tka, the 66 y/o female patient had a revision to replace a slope + insert due to excessive wear.The 9mm slope+ insert was replaced with a 13mm crc insert.There was no breakage of a device.There was no report of surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain x-rays.Photos received.The device is not available for evaluation as device is still at hospital and it has to go through risk management.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections d1, d4, g3, g4, g6, h2, h3, h4 and h6 have been updated accordingly.(h3) the revision reported was likely the result of axial rotational mismatch between the femoral and tibial components, third body wear, instability, patient-related conditions, or any combination of these possibilities, which led to polyethylene wear and delamination of the tibial insert.However, this cannot be confirmed as the devices were not returned for evaluation.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT SLOPE +, SZ 2.5, 9 MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14691495
Report Number1038671-2022-00687
Device Sequence Number5327342
Product Code JWH
UDI-Device Identifier10885862159182
UDI-Public10885862159182
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/20/2024
Device Model NumberLOGIC CR TIB INSERT SLOPE +, SZ 2.5, 9 MM
Device Catalogue Number02-012-48-2509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-