Model Number 050-87222
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the second stay safe connector during step 8 of their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Fluid was not discovered in the pump module area and on the external of the cassette.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Additional information was requested, however; to date was not provided.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the second stay safe connector during step 8 of their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Fluid was not discovered in the pump module area and on the external of the cassette.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Additional information was requested, however; to date was not provided.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY PDX, INTEGRATED SET
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14691698
Report Number0008030665-2022-00676
Device Sequence Number17458600
Product Code FKX
UDI-Device Identifier00840861101993
UDI-Public00840861101993
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K141145
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2022
Device Model Number050-87222
Device Catalogue Number050-87222
Device Lot Number21PR08156
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; DELFLEX PD FLUID; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER