| Model Number |
050-87222 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the second stay safe connector during step 8 of their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Fluid was not discovered in the pump module area and on the external of the cassette.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Additional information was requested, however; to date was not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the second stay safe connector during step 8 of their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Fluid was not discovered in the pump module area and on the external of the cassette.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Additional information was requested, however; to date was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|