NUVASIVE, INCORPORATED NUVASIVE COROENT SMALL INTERLOCK II SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL Back to Search Results
Model Number 6770815CP2
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 a patient underwent an anterior cervical interbody fusion procedure on unknown levels of the spine.During the procedure during implant insertion the cage cracked down the middle.It is unknown whether the cracked device remained in-situ or was removed from the patient.No injury or adverse effects to the patient were reported.
 
Additional Manufacturer Narrative
No product was returned and no radiographs or other information was provided confirming the alleged incident.Multiple attempts were made to obtain additional information from the involved reported with no response.Due to a complete lack of information the route cause of the event could not be determined however, excessive screw angulation, off centered placement and implant selection were identified as possible cause or contributors.No additional investigation can be completed.Label review: "warnings, cautions and precautions - correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant." "pre-operative warnings - care should be used in the handling and storage of the coroent implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the coroent implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the coroent implants if there is any evidence of damage.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "handling of the sterile implant - open the packages carefully.Take suitable measures to ensure that the implant does not come into contact with objects that could damage its surfaces.Use only the recommended instruments for implantation of the implants.Damaged implants must not be used.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE COROENT SMALL INTERLOCK II SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key14691770
Report Number2031966-2022-00098
Device Sequence Number5327347
Product Code OVE
UDI-Device Identifier00887517812452
UDI-Public887517812452
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K170961
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6770815CP2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Initial Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-