• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 723-00-36F
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Patient reported primary surgery on (b)(6) 2022 and reported 3 left hip dislocations with the most recent on (b)(6) 2022.Patient would like to know if parts are subject to a recall.Patient indicated he is in the process of retaining an attorney.(b)(6) 2022: 1st dislocation (approx), (b)(6) 2022: 2nd dislocation (approx), (b)(6) 2022: 3rd dislocation (approx), each dislocation was treated to "pop back in", nothing was removed.This is for dislocation reduction on (b)(6) 2022.
 
Additional Manufacturer Narrative
Reported event: an event regarding dislocation involving a trident liner was reported.The event was not confirmed.Method & results: -device evaluation and results: material analysis, visual, functional, and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were provided for review with clinical consultant.-device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been 3 other similar events for the reported lot.They relate to dislocation for the same device/patient, therefore no further trending is required for this commonality.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It is also noted that the patient is inquiring if the devices are part of a recall.Based on the catalog and lot codes provided the devices reported as not subject to a recall.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient reported primary surgery on (b)(6) 2022 and reported 3 left hip dislocations with the most recent on (b)(6) 2022.Patient would like to know if parts are subject to a recall.Patient indicated he is in the process of retaining an attorney.(b)(6) 2022: 1st dislocation (approx).(b)(6) 2022: 2nd dislocation (approx).(b)(6) 2022: 3rd dislocation (approx).Each dislocation was treated to "pop back in", nothing was removed.This is for dislocation reduction on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG X3 INSERT 36MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14691885
Report Number0002249697-2022-00832
Device Sequence Number6778477
Product Code LPH
UDI-Device Identifier07613327342239
UDI-Public07613327342239
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K182468
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number723-00-36F
Device Catalogue Number723-00-36F
Device Lot NumberMD0AD7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer08/08/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age57 YR
Patient SexMale
Patient Weight100 KG
-
-