Model Number 6366-9-220
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Revision (hmrs parts replacement after 12 years) according to bfarm special approval.Update 25may22, after 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
 
Additional Manufacturer Narrative
Reported event: an event regarding instability involving a hmrs axle was reported.The event was not confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that the device is damaged along the rim of the pin.This damage is likely due to contact with explantation tooling.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to worn poly components and joint instability.Visual inspection of the returned device indicated that the device is damaged along the rim of the pin.This damage is likely due to contact with explantation tooling.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revision (hmrs parts replacement after 12 years) according to bfarm special approval.Update (b)(6) 2022 after 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMRS AXIS PIN
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key14691914
Report Number0002249697-2022-00833
Device Sequence Number6637055
Product Code KRO
UDI-Device Identifier07613327049930
UDI-Public07613327049930
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6366-9-220
Device Catalogue Number63669220
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
-
-