• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ACU0T0
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during implantation of an intraocular lens (iol) the plunger went over the lens.There was no patient harm.The surgery was completed by a back up lens.
 
Additional Manufacturer Narrative
The device with the lens was returned loose in a plastic bag.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger has advanced to mid-nozzle and has overrode the lens.The lens was partially advanced into the nozzle.The trailing haptic was folded onto the anterior optic surface.The leading haptic was extended along the right side of the plunger.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.The root cause cannot be determined for the reported complaint.A plunger override was observed.It is unknown if a qualified viscoelastic was used.The ifu (instructions for use) instructs: during device preparation and implantation of the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Plunger override may occur: due to rapid advancement faster than the ifu recommend rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to override the lens.If the device is overfilled with ovd, this can prevent the trailing haptic from being placed properly or move the lens out of position resulting in misfolding.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.There are two other complaints in this lot.Investigation has been completed based on current information.Based on our current tracking, there are no adverse trends for this reported complaint.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14692468
Report Number1119421-2022-01274
Device Sequence Number6692565
Product Code KYB
UDI-Device Identifier00380652395335
UDI-Public00380652395335
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2024
Device Model NumberACU0T0
Device Lot Number15151855
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-