| Catalog Number |
07027117190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys cmv igg assay on a cobas 8000 e 801 module (serial number (b)(4)).The sample resulted in a cmv igg value of 0.31 u/ml (non-reactive) when tested on the e 801 analyzer.The value from the e 801 analyzer was reported outside of the laboratory.An aliquot of this sample was sent to another laboratory for testing using the diasorin liaison xl method, resulting in a value of 104 u/ml (immune/reactive).
|
| |
|
Additional Manufacturer Narrative
|
|
Samples from the patient were provided for investigation and the results obtained by the customer could be reproduced.
|
| |
|
Additional Manufacturer Narrative
|
|
For calibrations performed on (b)(6)2022 and (b)(6)2022, signals were within expected ranges.On (b)(6)2022, controls recovered within specified ranges.Further investigations of the patient sample determined it to be correctly non-reactive with the elecsys cmv igg assay.The non-reactive elecsys cmv igg result was supported by mikrogen recomline blot results, which were negative for cmv igg and reactive for cmv igm.Further investigation indicated that a primary infection was unlikely due to the constant igm response and the absence of a seroconversion.
|
| |
|