• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY KIT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY KIT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/870/075CZ
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use of the tracheal tube, the inflation line broke.No additional information was available.No patient injury was reported.
 
Additional Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).No problem or issues were identified during the device history record review.One picture and one sample were received for evaluation.One used decontaminated sample was returned for investigation without its original packaging.During visual inspection, we noticed that inflation line was detached from pilot balloon.Root cause was found to be a supplier item fault.This issue was escalated to a corrective action preventative action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY KIT
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key14695678
Report Number3012307300-2022-11932
Device Sequence Number20127711
Product Code BTO
UDI-Device Identifier15019517076219
UDI-Public15019517076219
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/870/075CZ
Device Catalogue Number101/870/075CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer04/25/2023
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-