A nurse reported that, during intraocular lens (iol) implant procedure, the plunger went over the iol.There was a contact with the patient but no patient harm involved.The surgery was delayed for a minute.The surgery was completed by using a back-up lens and without any other problem.Additional information was requested.
|
|
The device with the lens was returned loose in a plastic bag.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was located at mid-nozzle and was engaged to the trailing optic edge.The lens was folded in an acceptable position at the fill line.The trailing haptic was folded onto the anterior optic surface.The leading haptic was extended with the distal haptic tip.The reported plunger override was not observed.The plunger, lens, and haptics are in acceptable positions for advancement per the instructions for use (ifu).Viscoelastic information was not provided.It is unknown if a qualified product was used.The reported plunger override was not observed.No problems are observed with the returned device.The plunger, lens and haptic positions are acceptable.A straight leading haptic position was observed.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the ifu.The ifu instructs: during device preparation and implantation of the company lens with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.A straight leading haptic may occur: ¿ due to the normal folding variations as indicated the ifu diagrams.¿ if the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.¿ if there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.¿ due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.The manufacturer internal reference number is: (b)(4).
|