• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FEM NAR POR PS LT SZ 6; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FEM NAR POR PS LT SZ 6; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL Back to Search Results
Model Number 1504-11-126
Medical Device Problem Codes Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Codes Fall (1848); Inadequate Osseointegration (2646)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient fell and collapse the femoral component further into the distal medial femur causing the cementless femoral component to loosen.No harm to patient.Revised to crs cemented with cement stem.And a poly swap.No surgical delay.Doi: (b)(6) 2022, dor: (b)(6) 2022, affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary =no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot =the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEM NAR POR PS LT SZ 6
Common Device Name
ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14697324
Report Number1818910-2022-10938
Device Sequence Number6637186
Product Code MBH
UDI-Device Identifier10603295041993
UDI-Public10603295041993
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-11-126
Device Catalogue Number150411126
Device Lot Number9788638
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer07/01/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date07/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS RP INSRT SZ6 6MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-