SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX UNCUFFED CLEAR PVC SOUTH POLAR PREFORMED M; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/134/045
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, a part of the tube was found deformed or kinked.No patient injury.
 
Additional Manufacturer Narrative
H10: device evaluation: three photos were included in agile attachments for evaluation; photos showed a bit kink in curvature.One sample was returned for investigation.The returned sample was received in used condition without original packaging.The sample was visually inspected under normal conditions of illumination to detect any condition that could cause the failure mode reported.No discrepancies were found.The returned sample was inflated using a syringe and was submerged under water to detect any leak.No leakage was detected.No other analysis was performed.Based on the analysis conducted in the sample provided, kinked failure mode was confirmed.The failure condition was possibly due to packaging method not followed correctly (tube not placed properly on curving plate).Failure was traced to manufacturing.For corrective action an awareness notification was performed to personnel to explain the importance of adherence to manufacturing procedure.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX UNCUFFED CLEAR PVC SOUTH POLAR PREFORMED M
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key14699341
Report Number3012307300-2022-11975
Device Sequence Number6692768
Product Code BTR
UDI-Device Identifier15019315020865
UDI-Public15019315020865
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K931735
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model Number100/134/045
Device Catalogue Number100/134/045
Device Lot Number4062691
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-