• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. FENESTRATED, POLYAXIAL SCREW; SIZE Ø6.5X40 MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. FENESTRATED, POLYAXIAL SCREW; SIZE Ø6.5X40 MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number F2911-06540-G1
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an everest fenestrated screw fractured at s1 post-operatively.The physician indicates revision surgery is not required.
 
Additional Manufacturer Narrative
Devices remains implanted.
 
Additional Manufacturer Narrative
Product inspections could not be performed as the product is still implanted in the patient and is therefore unavailable for evaluation.However, an x-ray was provided and reviewed.The construct was a 2 level system with an interbody device.The screw implanted at s1 on the left side was highlighted in the x-ray.Although a clear fracture site cannot be observed, there does appear to be a slight bend in the shaft of the screw.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.The everest instructions for use states: "internal fixation devices are load sharing devices which maintain alignment until healing occurs.If healing is delayed or does not occur the implant could eventually break, bend, or loosen.Loads produced by load bearing and activity levels will impact the longevity of the implant.Metallic implants can loosen, fracture, corrode, migrate, cause pain, or stress shield bone even after a bone has healed.If an implant remains implanted after complete healing, it can actually increase the risk of re-fracture in an active individual.The surgeon should weigh the risks versus benefits when deciding whether to remove the implant.Implant removal should be followed by adequate postoperative management to avoid refracture." additional information regarding the patient's condition is unavailable at this time.As the screw has been implanted for approximately 1 year, the most likely root cause for the screw fracture can be attributed to material fatigue over time.
 
Event or Problem Description
It was reported that an everest fenestrated screw fractured at s1 post-operatively.The physician indicates revision surgery is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FENESTRATED, POLYAXIAL SCREW; SIZE Ø6.5X40 MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14700025
Report Number3004774118-2022-00237
Device Sequence Number5327043
Product Code NKB
UDI-Device Identifier10888857262195
UDI-Public10888857262195
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K161369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberF2911-06540-G1
Device Catalogue NumberF2911-06540-G1
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-