| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: norway.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial hip arthroplasty, the instrument fractured.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the tip of the instrument broke off at the threads, and a portion remained in the shell.The shell was not implanted.No further evaluation can be made from the provided pictures.Reported event was confirmed by review of photographs provided.Device history record (dhr) review was unable to be performed as the item and lot number of the device involved in the event is unknown.The reported failure has been previously investigated.As per the investigation, the failure mode occurs most frequently when the straight shell inserter is paired with the continuum or trilogy it shell.The levering force puts great stress on the distal end of the bolt and its interface with the shell threads.Initially, due to difference in the material properties, the titanium threads of the shell will yield before the stainless steel threads of bolt, but the shell is single use so that is acceptable.However, after repetitive cyclic loading, fatigue effects accumulate for the bolt and cause the tip to fracture.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|