• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AURIS HEALTH, INC. MONARCH PLATFORM; BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AURIS HEALTH, INC. MONARCH PLATFORM; BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES Back to Search Results
Model Number ASMITH6@ITS.JNJ.COM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 04/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a monarch-assisted bronchoscopy procedure due to concerns for potential pneumothorax (high peak pressure and declining 02 saturations).The physician obtained a portable chest x-ray (cxr) and a left sided pneumothorax was noted.The physician proceeded to place a 14g wayne pneumothorax chest tube at the bedside.The patient was admitted for observation.The chest tube was removed, and the patient was discharged the following day.
 
Additional Manufacturer Narrative
This event was reported from the transbronchial biopsy assisted by robot guidance in the evaluation of tumors of the lung (target) study (b)(6).The device was not returned for evaluation.There was no allegation of device failure.Risk of pneumothorax is documented in (b)(4), rev f, monarch bronch risk management report and is deemed to be acceptable.Based on the information available for this complaint, the root cause of the reported event is related to a known inherent risk of the device and procedure as documented in the labeling.
 
Additional Manufacturer Narrative
Engineering confirmed that the logs of the reported event were not being uploaded to the cloud.Three good faith efforts (gfe) were made acquiring the logs.However, the logs have not become available as of 7/18/2022.Due to the absence of logs, a full failure analysis cannot be performed against the system.A review of the device history record (dhr) was performed.There are no reports of nonconformance that relate to the reported incident.A review for similar complaint shows the reported issue is a known issue however the root cause is specific to each individual case.Risk of pneumothorax is documented in 105-000299-01, rev f, monarch bronch risk management report and is deemed to be acceptable.This issue will be tracked and trended in the monthly complaint trend review meetings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH PLATFORM
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES
Manufacturer (Section D)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer (Section G)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer Contact
andria smith
150 shoreline drive
redwood city, CA 94065
MDR Report Key14700731
Report Number3014447948-2022-00008
Device Sequence Number5175357
Product Code EOQ
UDI-Device IdentifierB634MON000005010
UDI-Public+B634MON000005010/16D20210922
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K193534
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberASMITH6@ITS.JNJ.COM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-