• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number 5388
Device Problems Display or Visual Feedback Problem (1184); Failure to Shut Off (2939); Inadequate User Interface (2958)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/31/2022
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis could not confirm customer comment that the epg) displayed an error code, or that the epg could not be turned off.Analysis confirmed the customer comment the epg setting could not be changed.The upper display was locked up.Rate, atrial, and ventricle numbers would not increase or decrease.The upper and lower case were broken.Two bail covers were broken.The ring cover was broken.Two bails were missing.The ring was missing.The keyboard was scratched.The liquid crystal display (lcd) was missing segments.The device was not repaired as it was at end of service.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) displayed an error code.It was also reported that the settings could not be changed or shut off.The epg returned into service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer (Section G)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14700936
MDR Text Key296862565
Report Number2183613-2022-00074
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P820003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5388
Device Catalogue Number5388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-