• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TI LOCKING SCREW; POSTERIOR CERVICAL SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TI LOCKING SCREW; POSTERIOR CERVICAL SCREW SYSTEM Back to Search Results
Model Number 04.614.508
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Drug Resistant Bacterial Infection (4553); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2022, the patient underwent revision surgery due to a migrated rod.The surgeon commented the following: the setscrews on both ends of the plate had come off.The rod was about to protrude from the occipital region.The surgeon confirmed that final screw tightening during the primary procedure was performed.The patient underwent primary posterior cervical fusion in the occipital bone for treating vestibular neuronitis on (b)(6) 2021.The procedure was completed without surgical delay.On (b)(6) 2022, the surgeon reported that the rod might have come off.This report is for a titanium (ti) locking screw.This is report 1 of 3 for (b)(4).Additional reports are captured under (b)(4).
 
Additional Manufacturer Narrative
Additional narrative: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Date of concomitant therapy is (b)(6) 2021.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: sterile part: part #: 04.614.508s.Lot #: 7l26613.Manufacturing site: werk selzach.Supplier: (b)(4).Release to warehouse date: 03 sep 2020.Expiration date: 01 jan 2030.Non-sterile part: part #: 04.614.508.Lot #: 32p5695.Manufacturing site: werk selzach.Supplier: (b)(4).Release to warehouse date: 06 jan 2020.Expiration date: n/a.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the ti locking screw shows markings consistent with the implantation and explanation of the device.No other irregularity or defects were observed on the returned device.No post-operative images to assess the migration condition were provided.A dimensional inspection was performed for the ti locking screw and met specifications.Major thread diameter and height were inspected using micrometer.The minor thread diameter was inspected using zz plus vermont gauge pin gage set.A functional test was not conducted since the mating device was not returned for examination.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the ti locking screw was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Although there is no loosening of implants, mrsa infection is confirmed.The surgeon commented that the patient's son is on the verge of suing for lack of explanation from the manufacturer about implant problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI LOCKING SCREW
Common Device Name
POSTERIOR CERVICAL SCREW SYSTEM
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14701200
Report Number8030965-2022-04007
Device Sequence Number14012052
Product Code NKG
UDI-Device Identifier10705034739229
UDI-Public(01)10705034739229
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142838
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.614.508
Device Catalogue Number04.614.508
Device Lot Number32P5695
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer06/20/2022
02/20/2023
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date07/18/2022
03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OCCIPITAL-PL 4.5/5 MED W/50 F/R Ø4 TI; TI LOCKING SCREW - STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-