| Model Number |
IPN920517 |
| Medical Device Problem Codes |
Leak/Splash (1354); Fitting Problem (2183)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Customer reported prior issue of "green seal is too large.When the hub is tightened, the seal is pushed over the thread and leads to leakage (air is drawn during aspiration)." no patient information was available.It was reported in each case the issue could be solved by using a replacement hub.It was unknown how many times the issue occurred.
|
| |
|
Event or Problem Description
|
|
Customer reported prior issue of "green seal is too large.When the hub is tightened, the seal is pushed over the thread and leads to leakage (air is drawn during aspiration)." no patient information was available.It was reported in each case the issue could be solved by using a replacement hub.It was unknown how many times the issue occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|